Position:home > News > New Product Display of Fangyuan Instruments: FFZ432 Retroreflective Coefficient Tester

New Product Display of Fangyuan Instruments: FFZ432 Retroreflective Coefficient Tester

    With the development of society and the improvement of living standards, cars as a means of transportation are becoming more and more popular, and the number of cars is increasing sharply. As a result, people who often need to cross the highway or work on the highway for a long time, such as cleaners and traffic police, are facing increasing risks.  In order to effectively reduce the occurrence of traffic accidents, it has become imperative to wear warning clothing made of special materials to such people, or to add special warning signs to people's clothes, roadblocks and warning signs.  In this way, whether it is under harsh light during the day or under the irradiation of dim lights at night, the driver can be sure to find pedestrians or roadblocks and warning signs in time, and the driver will have enough time to brake or turn to avoid accidents.  Whether warning clothing or warning signs can be found in time in a specific environment or at a specific distance becomes a crucial factor to avoid traffic accidents.  "Retroreflective Coefficient Tester" simulates whether the light intensity fed back from the vehicle driving at night under the specified standard light source, distance and angle meets the requirements of GB206532006, GB/T284682012 and other standards, thus effectively evaluating and judging the warning signs.  FFZ432 retroreflective coefficient tester produced by Wenzhou Fangyuan Instrument Co., Ltd. is divided into manual measurement and automatic measurement. Manual measurement can be used to quickly measure a certain incident angle and observation angle.  The automatic test can carry out comprehensive measurement for 16 angles required by the standard, does not need human participation and operation after starting, and is simple and quick to test.

 

 

    FFZ432 retroreflective coefficient tester

    Uses: Used to evaluate the reflective performance of high visibility warning clothing, traffic safety reflective school uniforms, warning signs and other warning materials in dry and wet conditions, and to measure the retroreflective coefficient of the sample surface.  Compliance with standards: GB 206532006 Occupational High Visibility Warning Clothing Appendix C Appendix D, GBT28468 and other standards.

 

    Instrument characteristics:

    1. Low - light - level photometric acquisition system is adopted, with high signal resolution and perfect data.

    2. Infrared ranging, quickly and accurately determine the relative position of the measurement and control table and the sample table.

    3. Wireless transceiver module is adopted, which makes instrument control and data docking at both ends faster and more convenient.

    4. Two modes of measurement and scanning measurement can be selected in the software to facilitate data analysis.

    5. Precision motor control drive, fast and accurate positioning, low noise.

    6. The incident angle and observation angle are automatically adjusted, and the test is simple and convenient.

    7. The light source system is stable and reliable with uniform vertical illuminance.

    8. The pressure stabilizing device makes the spray pressure uniform and easy to control, and is equipped with a water collecting tank and a collector.

    9. Quick - change structure is adopted, and sample clamping is simple and quick.

    10. The core control components are Italy's 32 - bit multi-functional motherboard.

    11. Imported special aluminum wire drawing panel with metal keys.

    12. Color touch screen display, menu mode of operation.

    13. Data output automatic printing.

    14. With online software functions.

 

    Technical parameters:

    1. Light source: standard light source A (color temperature 2856K).

    2. Distance between photodetector and sample surface: 15m adjustable (resolution 1mm).

    3. Measuring range of light detector:

        3.1. Range of main console sensor: 0.0000120Lx.

        3.2. Measurement range of sample table sensor: 0.1 ~ 300 LX.

    4. The observation angleα12′2°can be adjusted.

    5. The incident angleβ140° can be adjusted.

    6. Vertical illuminance uniformity 4 %.

    7. Numerical reproducibility of retroreflective coefficient: ≤±3% (standard darkroom).

    8. The nozzle is 1000±2mm from the sample.

    9. Velocity of sprinkler precipitation per hour: 284 mm / H.

    10. Power supply: 220V, 50Hz, 1kW.

    11. Overall dimensions of the main console: 900mm long, 780mm wide and 1700mm high.

    12. Overall dimensions of sample table: 630mm long, 430mm wide and 1900mm high.

    13. Weight of sample table: 80kg.

    14. Main console weight: 115kg.