FXC852 Ozone resistance tester

INTRODUCTION

APPLICATION: 

It can be used for aging test of nonmetallic materials and organic materials (coatings, rubber, plastics, paints, pigments, etc.) under ozone condition.

STANDARDS: 

GB/T7762, ASTM1149, ISO1431, GB/T13642, HG/T2869, etc.

Instrument characteristics:

1. Large - screen color touch screen display, user-friendly interface design and easy operation.

2. The panel is made of special aluminum and has metal keys, which are sensitive to operation and not easy to damage.

3. The core control components are Italy's 32 - bit multi-functional motherboard.

4. The core transmission mechanism adopts imported precision guide rails.

5. The instrument includes Chinese and English interfaces to facilitate the inspection of foreign customers.

6. The surface of the instrument is treated by electrostatic spraying process.

Technical parameters:

1. Studio size (mm): 400× 400× 500

2. Ozone concentration: 25 ~ 1000 pphm. (adjustable)

3. Ozone concentration deviation: ≤ 5 %

4. Laboratory temperature: RT + 10 ℃ ~ 60 ℃

5. Temperature fluctuation: ±0.5℃

6. Uniformity: ±2℃

7. Test gas flow rate: 20 ~ 80L / min

8. Test device: static

9. Sample rack rotating speed: 360 rotating sample rack (rotating speed 1 rpm) static sample rack

10. Source of ozone: ozone generator (ozone generated by voltage silent discharge tube)

11. Sensors: ozone concentration sensors imported from Britain can realize accurate control.

12. The controller adopts Panasonic PLC of Japan (Panasonic PLC imported from Japan)